DATA SHEET BC4 single battery analyser |
![]() |
|||||
Updated: 02 Feb 2009 |
||||||
Description: |
Battery analyser/conditioner/recharger | |||||
Product name: |
BC4 (BC4-C if fitted with optional serial port) |
|||||
Use: |
The BC4 single battery analyser makes it possible not only to recharge batteries used in the SE4 family of respirators, but also to ascertain the current status of the battery, as well as maintaining the battery in peak condition over long periods of time. |
|||||
Supplied AC power: |
100–240 V, 50–60 Hz |
|||||
DC power (charging mode, max) |
21 V, 2.5 A |
|||||
| Charging current: |
|
|||||
| Post-trickle charging current: | 0.2 A |
|||||
| Pre-trickle recovery current: | 0.75 A |
|||||
| Max critical battery temperature: | +62°C (144°F) |
|||||
| Max critical PCB temperature: | +95°C (203°F) | |||||
| Max allowed charging start temperature: | +32°C (90°F) | |||||
| Min allowed charging start temperature: | +5°C (41°F) | |||||
| Post-trickle charging duration: |
|
|||||
Resistance PASS: |
|
|||||
Resistance FAIL critical (charging): |
>800 mohm | |||||
| Maximum allowed voltage during charging: | 17 V | |||||
| Minimum allowed voltage during charging: | 8 V |
|||||
| Discharge level: | 10.5 V | |||||
| Discharge current: | 1 A | |||||
Emergency cut-off current: |
|
|||||
| Pre-programmed charging/discharging cycles: | 0–5 | |||||